SEA-BIRD ELECTRONICS, INC.

                          1808 136th Place N.E., Bellevue,   Washington 98005  USA

                               Phone: (425) 643 - 9866 Fax: (425) 643 - 9954     Internet: seabird@seabird.com

 

SENSOR SERIAL NUMBER = 0020                    SBE 43

CALIBRATION DATE: 13-Nov-01repair           OXYGEN CALIBRATION DATA

 

COEFFICIENTS:

Soc  =  0.4281                        TCor =  0.0007

Boc  =  0.0127                         PCor =  1.350e-04

Voffset =  -0.6119

 

BATH OX      BATH TEMP      BATH SAL         INSTRUMENT        INST OX         RESIDUAL

ml/l                  (ITS-90 degsC)     PSU                       VOLTS                   ml/l                   ml/l

1.38                                 25.00                0.06                                 1.148                              1.38                      -0.00

1.52                                   5.00                0.07                                 0.984                              1.52                       0.00

2.38                                 25.00                0.06                                 1.542                              2.37                      -0.00

2.52                                   5.00                0.07                                 1.247                              2.53                       0.01

3.37                                 25.00                0.06                                 1.940                              3.37                      -0.00

3.57                                   5.00                0.07                                 1.518                              3.57                       0.00

5.40                                 25.00                0.06                                 2.748                              5.40                       0.00

5.59                                   5.00                0.07                                 2.044                              5.59                      -0.00

 

7.33                                  5.04                0.07                                 2.501                              7.33                      - 0.00

 

Note: Replaced the electrolyte and oxygen membrane

         Re-backfilled the sensor’s electrolyte reservoir

         De-gassed the sensor’s electrolyte solution

         Installed new style DO plenum

         Recal’d the sensor

        Performed full diagnostic evals