SEA-BIRD ELECTRONICS, INC.

                          1808 136th Place N.E., Bellevue,   Washington 98005  USA

                               Phone: (425) 643 - 9866 Fax: (425) 643 - 9954     Internet: seabird@seabird.com

 

SENSOR SERIAL NUMBER = 0020                    SBE 43

CALIBRATION DATE: 01-May-03                      OXYGEN CALIBRATION DATA

 

COEFFICIENTS:

Soc  =  0.4630                        TCor =  0.0010

Boc  =  0.0000                         PCor =  1.350e-04

Voffset =  -0.5940

 

BATH OX      BATH TEMP      BATH SAL         INSTRUMENT        INST OX         RESIDUAL

ml/l                  (ITS-90 degsC)     PSU                       VOLTS                   ml/l                   ml/l

1.68                                 25.00                0.03                                 1.206                              1.67                      - 0.00

1.83                                   5.00                0.04                                 1.034                              1.83                        0.00

2.50                                 25.00                0.03                                 1.508                              2.50                        0.00

2.63                                   5.00                0.04                                 1.224                              2.62                      - 0.01

3.35                                 25.00                0.03                                 1.818                              3.35                        0.00

3.42                                   5.00                0.04                                 1.417                              3.42                       0.00

5.08                                 25.00                0.03                                 2.451                              5.09                       0.01

5.18                                   5.00                0.04                                 1.843                              5.19                       0.01

6.67                                   5.00                0.04                                 2.200                              6.67                      -0.00

6.88                                 25.00                0.03                                 3.103                              6.87                      -0.01

 

 

Note: Post calibrated the sensor .  Backfilled the sensors pressure compensation bag with saline solution

Performed full diagnostic eval, there were no problems found.

Note: problem described by customer may have been caused by a bad cable.