SEA-BIRD ELECTRONICS, INC.
Phone: (425) 643 - 9866 Fax:
(425) 643 - 9954 Internet:
seabird@seabird.com
SENSOR
SERIAL NUMBER = 0019 SBE 43
CALIBRATION
DATE: 26-Sep-02p OXYGEN CALIBRATION DATA
COEFFICIENTS:
Soc = 0.3288 TCor
= 0.0006
Boc = 0.0000 PCor
= 1.350e-04
Voffset
= -0.5844
BATH OX BATH TEMP BATH SAL INSTRUMENT INST OX RESIDUAL
ml/l (ITS-90 degsC) PSU VOLTS ml/l ml/l
1.46 5.00 0.04 1.077 1.45 -0.01
1.54 25.00 0.03 1.387 1.54 0.01
2.44 5.00 0.04 1.412 2.44 -0.01
2.58 25.00 0.03 1.929 2.59 0.01
3.50 5.00 0.04 1.772 3.50 -0.00
3.63 25.00 0.03 2.474 3.64 0.01
5.54 5.00 0.04 2.469 5.55 0.00
5.75 25.00 0.03 3.580 5.76 0.01
7.12 5.00 0.04
3.004 7.12 0.00
7.99 25.00 0.03 4.728 7.97 -0.01
Sensor was
found to have flood damage.
Note:
Calibrations and services performed on SBE 43 DO sensor
Replaced the electronics & module assembly
Recalibrated the sensor
Performed full diagnostic evaluation