SEA-BIRD ELECTRONICS, INC.

                          1808 136th Place N.E., Bellevue,   Washington 98005  USA

                               Phone: (425) 643 - 9866 Fax: (425) 643 - 9954     Internet: seabird@seabird.com

 

SENSOR SERIAL NUMBER = 0019                    SBE 43

CALIBRATION DATE: 09-Jan-02w                     OXYGEN CALIBRATION DATA

 

COEFFICIENTS:

Soc  =  0.3461                         TCor =  0.0000

Boc  =  -0.0476                         PCor =  1.350e-04

Voffset =  -0.6112

 

BATH OX      BATH TEMP      BATH SAL         INSTRUMENT        INST OX         RESIDUAL

ml/l                  (ITS-90 degsC)     PSU                       VOLTS                   ml/l                   ml/l

1.31                      25.00                0.03                            1.332                        1.31                      -0.00

1.35                        5.00                0.04                            1.151                        1.30                      -0.05

2.16                      25.00                0.03                            1.762                        2.17                       0.01

2.20                        5.00                0.04                            1.439                        2.19                      -0.01

2.99                      25.00                0.03                            2.184                        3.01                       0.02

3.10                        5.00                0.04                            1.733                        3.10                      -0.00

4.73                      25.00                0.03                            3.049                        4.74                       0.01

4.91                        5.00                0.04                            2.326                        4.93                       0.02

6.33                      25.00                0.03                            3.839                         6.31                      -0.02

7.43                        5.00                0.04                            3.128                         7.41                      -0.02

 

      Note: Calibrations and services performed on SBE 43 DO sensor

               Replaced the electrolyte and oxygen membrane

              Installed a new style DO plenum

             Recalibrated the sensor