Conductivity Sensor Calibration Sheet

                             
  SEA-BIRD ELECTRONICS, INC.    
  1808 136th Place N.E., Bellevue, Washington 98005 USA    
  Phone: (425) 643 - 9866 Fax: (425) 643 - 9954 Internet: seabird@seabird.com    
                             
  SENSOR SERIAL NUMBER: 1176   CONDUCTIVITY CALIBRATION DATA          
  CALIBRATION DATE: 02-Aug-01s   PSS 1978: C(35,15,0)=4.2914 Siemens/meter        
                             
                             
  GHIJ COEFFICIENTS   ABCDM COEFFICIENTS              
  g= -4.31707525E+00   a= 9.81223105E-05   M= 3.6        
  h= 5.93676113E-01   b= 5.93933142E-01   CPcor= -9.57E-08        
  i= 1.93633266E-04   c= -4.31736821E+00   CTcor= 3.25E-06        
  j= 2.24893780E-05   d= -8.07850481E-05              
                             
                             
                             
                             
  BATH TEMP BATH SAL BATH COND INST FREQ INST COND RESIDUAL    
    (IPTS-68 degsC) (PSU) (Siemens/m) (kHz) (Siemens/m) (Siemens/m)    
    0.0000 0.0000 0.00000 2.69507 0.00000 0.00000    
    -1.3571 36.0219 2.86135 7.43094 2.86132 -0.00003    
    1.1123 36.0216 3.07895 7.67181 3.07899 0.00004    
    15.3474 36.0235 4.43838 9.03021 4.43840 0.00002    
    18.8516 36.0249 4.79687 9.35512 4.79682 -0.00005    
    29.3692 36.0249 5.91705 10.30371 5.91710 0.00005    
    32.9753 36.0226 6.31381 10.61892 6.31378 -0.00003