SEA-BIRD ELECTRONICS, INC.
Phone: (425) 643 - 9866 Fax:
(425) 643 - 9954 Internet:
seabird@seabird.com
SENSOR
SERIAL NUMBER = 1176 CONDUCTIVITY CALIBRATION DATA
CALIBRATION
DATE:
GHIJ
COEFFICIENTS ABCDM COEFFICIENTS
g = -4.31921063e+00 a = 5.79390797e-05
h =
5.94248029e-01 b
= 5.94282636e-01
i = 6.69909348e-05 c = -4.31891661e+00
j =
3.05021127e-05 d
= -7.86757782e-05
CPcor = -9.57e-08
(nominal) M
= 3.8
CTcor = 3.25e-06
(nominal)
CPcor = -9.57e-08 (nominal)
BATH
TEMP BATH SAL BATH COND INST FREQ INST COND RESIDUAL
(IPTS-68
degsC)
(PSU)
(Siemens/m) (kHz)
(Siemens/m) (Siemens/m)
0.0000 0.0000 0.00000 2.69508 -0.00000 -0.00000
-0.5001 35.1408 2.87109 7.44172 2.87110 0.00001
1.0000 35.1414 3.00109 7.58627 3.00108 -0.00001
14.9999 35.1424 4.30735 8.90802 4.30734 -0.00001
18.4999 35.1426 4.65694 9.22925 4.65695 0.00001
29.0000 35.1399 5.74932
10.16663 5.74933 0.00001
32.5000 35.1345 6.12516 10.46935 6.12515 -0.00001
POST CRUISE CALIBRATION