HOT-244 Optics Readme File by: Benedetto Barone Instruments Serial DevFile LastCalibration ----------------------------------------------------------- ACS 040 acs040.dev 7/6/12 AC9 281 ac90281.dev 3/10/11 CTD SBE37 SBE37.dev 3/31/05 ECO FL3 654 FL3-654.dev 12/14/06 LISST 1421 factory_zsc_1421.asc FRRf …82014 Licor profiler 046063 Pump 4.5K (on ACS) Pump 3K (on AC9) ----------------------------------------------------------- Hperpro2 120 Licor deck - ----------------------------------------------------------- 7/30/12 -HYPERPRO DECK: clean, start logging underway -LICOR DECK: clean, start logging -DH4: reset time -HYPERPRO: dark count h244_dark.raw -HYPERPRO Kahe cast 1240 - lat: 21 deg 20.591'N - lon: 158 deg 16.394'W - clouds: 10% - swell: 0.5 m 7/31/12 -HYPERPRO Aloha cast 1330 - lat: 22 deg 45.017'N - lon: 158 deg 0.003'W - clouds: 40% - swell: 1 m -AC9 air calibration: ac9_073112_air.dat -AC9 water calibration: ac9_073112_26_5.dat air water c488 0.472 0.135 a555 0.056 0.137 c650 0.176 0.142 a488 0.094 0.141 c555 0.408 0.141 a676 -0.263 0.123 T = 26.5 C -ACS air calibration: acs_073112_air.dat -ACS water calibration: acs_073112_26_5.dat air water a661 -0.395 0.000 a570 -0.110 0.001 a480 -0.066 0.005 c660 -0.133 0.006 c569 0.181 0.004 c480 0.250 0.002 T = 26.5 C -ECO FL3: dark count h244cal1.raw, reset time -FRRF: format memory, charge battery -BATTERY PACKS: charged, 13.0 & 13.0 V 8/1/2012 -LISST: battery 7.79 V, reset time, calibration h244_L1241_080112_cal.asc -DH4: clear memory 8/2/2012 AC9/ACS/FRRF/LISST/ECO3 cast: - in:303, down:309, bottom:328, up:327, out:351 (filters on AC9/ACS) - in:356, down:403, bottom:422, up:423, out:446 (filters on AC9) - depth : ~ 190m (probably shallower due to wire angle) - rate : 10 m/min down/up - weather : night , 13 knots winds E - location : 22° 45.038' N, 158° 00.016' W -LISST: battery 7.66 V -FRRF: download files, format memory -BATTERY PACKS: charge ---------- -PROBLEM: pin of the charger broke into the bulkhead connector of one of the two battery packs. Both cable of the charger and bulkhead connector need substitution ---------- AC9/ACS/FRRF/LISST/ECO3 cast: - in:1150, down:1156, bottom:1215, up:1216, out:1239 (filters on AC9/ACS) - in:1243, down:1249, bottom:1307, up:1308, out:1331 (filters on AC9) - depth : ~ 190m (probably shallower due to wire angle) - rate : 10 m/min down/up - weather : 10% clouds , 14 knots winds ESE, 1' swells - location : 22° 44.157' N, 158° 01.892' W Comments: could temporarily connect battery with broken bulkhead, sealed with grease. Apparently DH4 recorder during the 2 casts. -HYPERPRO Aloha cast 1340 - lat: - lon: - clouds: 25% - swell: 1 m -BATTERY PACKS: charged one per time -LISST: download files, disconnect battery, clean pressure sensor cavity, clean windows -ACS/AC9: clean windows and flow tubes -FRRF: download files, format memory, charge battery -DH4: download files, clear memory -LICOR: download data -ISUS: download data -FRRF downloaded files: pro1h244.dat pro2h244.dat pro3h244.dat pro4h244.dat -DH4 downloaded files: run.000 run.001 run.002 run.003 -LISST 1421 downloaded files: L2150300.dat L2150354.dat L2151148.dat L2151241.dat -LICOR downloaded file: licor_h244.txt Comment: it only started logging on 7/31 because of settings used during Hoe-Dylan -HTPERPRO downloaded files: h244_dark.raw h244_uw1.raw, h244_uw2.raw, h244_uw3.raw, h244_uw4.raw h244_kahe1.raw, h244_kahe2.raw, h244_kahe3.raw h244_aloha1.raw, h244_aloha2.raw, h244_aloha3.raw h244_aloha4.raw, h244_aloha5.raw