HOT-242 Optics Readme File by: Benedetto Barone, Dan Sadler 5/22/12 HOTSPOT for calibration and pre-cruise prep -Problems with the pressure sensor of LISST, high pressure out of water caused by salt accumulated in the cavity of the pressure sensor. Problem still present after flushing. -Built 2 new battery packs, installed with Y cable. -Cleaned ACS windows and flow tubes -ACS air calibration: H242_052212_air.dat -ACS water calibration: H242_052212_29_8.dat 5/29/12 -FRRF: charged, formatted memory -BATTERY PACKS: charged -DH4: reset time, cleared memory -AC9 air calibration: AC9_052912_air.dat -AC9 water calibration: AC9_052912_26_1.dat air water c412 0.445 0.204 c488 0.321 0.047 c555 0.258 0.015 c715 -0.563 0.064 a412 0.091 0.071 a488 -0.024 0.021 a555 -0.065 0.015 a715 -0.912 0.018 -ACS air calibration: acs_052912_air.dat -ACS water calibration: acs_052912_31_8.dat air water a659.0 -0.368 0.008 a567.6 -0.087 0.008 a480.3 -0.053 0.004 c662.4 -0.128 0.050 c570.7 0.191 0.055 c479.0 -0.260 0.043 -ECO3 calibration: h242ecocal1.raw -LISST calibration: h242_052912_cal1.asc, h242_052912_cal2.asc, h242_052912_cal3.asc 5/30/12 AC9/ACS/FRRF/LISST/ECO3 cast: - in:0301, down:0309, up:0329, out:0350 (filters on AC9/ACS) - in:0400, down:0408, up:0428, out:0453 (filters on AC9) - depth : ~ 190m (probably shallower due to wire angle) - rate : 10 m/min down/up - weather : night, 22 knots winds, 10' swells - location : 22° 44.666' N, 157° 59.855' W -Cage banged ship on first recovery -All lights blinking at the end of the 2 casts -LISST pressure not working (check on file) -FRRF charged -BATTERY PACKS: charged AC9/ACS/FRRF/LISST/ECO3 cast: - in:1320, down:1329, up:1347, out:1410 (filters on AC9/ACS) - in:1416, down:1424, up:1444, out:1509 (filters on AC9) - depth : ~ 190m (probably shallower due to wire angle) - rate : 10 m/min down/up - weather : 40% clouds , 17 knots winds, 7' swells - location : 22° 45.894' N, 158° 00.008' W 5/31/12 -ACS air calibration: acs_053112_air.dat -ACS water calibration: acs_053112_23_5.dat -ACS water calibration only c tube: acs_053112_23_7_onlyc.dat air water a659.0 -0.378 0.007 a567.6 -0.095 0.007 a480.3 -0.062 0.006 c662.4 -0.106 0.048 c570.7 0.212 0.055 c479.0 0.274 0.039 -LISST disconnected battery -FRRF downloaded files: pro1h242.dat pro2h242.dat pro3h242.dat pro4h242.dat -DH4 downloaded files: run.001 9144 Kb run.002 7329 Kb run.003 2977 Kb (anomalous small file) run.004 9356 Kb -LISST downloaded files: L1510256.dat L1510355.dat L1511317.dat L1511412.dat