HOT-200 Optics Readme file 2/21/08 Charging MPAK and FRRF batteries 2/22/08 Checking MPAK : 1330 - voltage = 12.1V - reset time * Unable to reformat flashcard! Attempting use of Host program. * SUCCESS! * ready to calibrate 2/23/08 Checking FRRF : 1410 - time OK - voltage = 14.45V - reformatted flashcard Cleaned AC-9 windows : 1315 Calibrating AC-9 - temp = 21.1 - files : run.000, 23feb08a1.dat, 23feb08c1.dat 2/24/08 FRRF/MPAK cast - in:1254, down:1300, up:1325, out:1353 - depth : ~250m using ship's capstan - rate : ~10 m/min down/up - weather : 8/8 clouds, 4-6' swells - location : 22 44.446 N, 158 0.873 W - files : run.002, pro1h200.bin 2/25/08 FRRF/MPAK cast - in:343, down:351, up:415, out:444 - depth : ~250m using ship's capstan - rate : ~10 m/min down/up - weather : night, 2-4' swells, 17 knott winds from 305 deg - location : 22 44.925 N, 157 59.916 W - files : run.003, pro2h200.bin Recharging MPAK : 458 Checking FRRF : 502 - voltage = 14.23 V - time OK - downloading : pro1h200.bin, pro2h200.bin - reformatting flashcard Checking MPAK : 1040 - voltage = 12.4V - reset time FRRF/MPAK cast - in:1124, down:1131, up:1156, out:1224 - depth : ~250m using ship's capstan - rate : ~10 m/min down/up - weather : 0/8 clouds, 12 knott winds from 325 deg, 4-6' swells - location : 22 41.232 N, 157 56.031 W - files : run.005, pro3h200.bin FRRF/MPAK Filter cast - in:1229, down:1234, up:1300, out:1326 - depth : ~250m using ship's capstan - rate : ~10 m/min down/up - weather : 3/8 clouds, 4-6' swells - location : 22 41.231 N, 157 56.032 W - files : run.006, pro4h200.bin Checking FRRF : 1436 - voltage = 14.03 V - time OK - downloading : pro3h200.bin, pro4h200.bin Checking MPAK : 1529 - voltage = 11.5 V - downloading : run.000, run.002, run.003, run.005, run.006 NOTES: I was unable to format the MPAK flashcard though the MPAK3 software, but I was able to by accessing the MPAK directly though hyperterminal.