HOT-181 Optics Readme file 5/24/06 Charging MPAK battery : ~1200 - note: using old charger 5/25/06 Checking MPAK : - voltage = 12.7V @ 921 - voltage = 13.0V @ 1027 - voltage = 12.4V @ 1141 Charging FRRF battery : 900 5/26/06 Checking MPAK : 631 - voltage = 112.6V - reset time - re-formating flashcard * Unable to reformat! The red indicator light stays on and I loose communication with the MPAK. I'm able to restart by disconnecting the battery and reconnecting. But, checking the flashdrive indicates that the files were not erased. Disconnecting for a few minutes. Still unable to reformat! Disconnecting for an hour. Finally able to reformat after multiple tries. - ready to calibrate Checking FRRF : 638 - voltage = 14.63V - time OK - re-formatted flashcard 5/27/06 Cleaned AC-9 windows : 910 Calibrating AC-9 : 925 - temp = 21.4V - calibrated A (1x) and C (1x) channels - files: run.000, 06may27a1.dat, 06may27c1.dat MPAK/FRRF cast - in:1258, down:1306, up:1333, out:1358 - depth : ~250m using ship's trawl winch - rate : ~10 m/min down/up - weather : 1/8 high clouds, 2-4' waves, 8 kt winds from 100 deg - location : 22 44.676 N, 157 59.466 W - files : run.001, pro1h181.bin * note : MPAK light still "blinking" when brought on deck Charging MPAK battery 5/28/06 Checking MPAK : 250 - voltage = 12.8V - reset time MPAK/FRRF night cast - in:305, down:311, up:337, out:408 - depth : ~250m using ship's capstan - rate : ~10 m/min down/up - weather : no stars, 2-4' swells, 13 knott winds from 100 Deg - location : 22 45.028 N, 157 59.985 W - files : run.003, pro2h181.bin * note : MPAK light still "blinking" when brought on deck Charging MPAK : 423 - downloading : run.000, run.001, run.003 - recharging Checking FRRF : 452 - time OK - voltage = 14.38V - downloading data : pro1h181.bin, pro2h181.bin - space used = 5446730, space left = 5039030 - reformatting flashcard Checking MPAK : 1047 - voltage = 13.1Y - reset time MPAK/FRRF cast - in:1132, down:1138, up:1204, out:1233 - depth : ~250m using ship's capstan - rate : ~10 m/min down/up - weather : 7/8 hazy clouds, 2-4' swells, 11 kt winds from 80 deg - location : 22 46.336 N, 157 55.678 W - files : run.005, pro3h181.bin * note : MPAK light still "blinking" when brought on deck MPAK/FRRF filter cast - in:1235, down:1243, up:1308, out:1335 - depth : ~250m using ship's capstan - rate : ~10 m/min down/up - weather : 7/8 hazy clouds, 4' swells, 11 kt winds from 80 deg - location : 22 46.336 N, 157 55.678 W - files : run.006, pro3h181.bin * note : MPAK wasn't "blinking" when brought on deck Checking FRRF : 1428 - time OK - voltage = 14.16V - downloading data : pro3h181.bin Checking MPAK : 1517 - downloading data : run.005, run.006 * Initially had problems communicating with instrument. Error condition when connecting battery. Worked fine after the 3rd try. * The battery discharged during the end of the downcast of the 2nd profile. NOTE: Using the old charger helped a little. The MPAK battery managed to last at least one profile. It was sent back to WetLabs to have it's core replaced.