HOT-180 Optics Readme file 3/31/06 Charging FRRF & MPAK batteries : 930 4/1/06 Checking FRRF : 625 - time OK - voltage = 14.72V - re-formatted flashcard Checking MPAK : 630 - reset time - voltage = 12.8V - re-formatted flashcard - ready to calibrate Calibrating AC-9 : 640 - cleaned AC-9 windows - temp = 23.1V - calibrated A (2x) and C (1x) channels - files: run.000, 06apr01a1.dat, 06apr01a2.dat, 06apr01c1.dat * Had trouble getting out all the bubbles. Recommend changing O-rings. 4/2/06 MPAK/FRRF cast - in:1307, down:1315, up:1339, out:1406 - depth : ~250m using ship's trawl winch - rate : ~10 m/min down/up - weather : 7/8 clouds, 3-4' waves, 20 kt winds from 85 deg - location : 22 47.657 N, 158 2.644 W - files : run.001, pro1h180.bin * note : MPAK wasn't "blinking" when brought on deck Charging MPAK battery : 1424 - 1737 Checking MPAK : 1738 - voltage = 12.6V - downloading data : run.000, run.001 * The battery appears to have suddenly discharged near the end of the upcast. - connecting to charger again 4/3/06 Checking MPAK : 239 - voltage = 13.0V MPAK/FRRF night cast - in:349, down:357, up:422, out:445 - depth : ~250m using ship's trawl winch - rate : ~10 m/min down/up - weather : 4-6' swells, 20 knott winds from 100 Deg - location : 22 45.031 N, 158 0.055 W - files : run.004, pro2h180.bin * note : MPAK wasn't "blinking" when brought on deck Charging MPAK : 450 - 1145 Checking FRRF : 501 - time OK - voltage = 14.37V - downloading data : pro1h180.bin, pro2h180.bin Checking MPAK : 1145 - downloading data : run.004 * The battery appears to have suddenly discharged during the upcast. MPAK/FRRF filter cast - in:1355, down:1342, up:1407, out:1433 - depth : ~250m using ship's trawl winch - rate : ~10 m/min down/up - weather : 4/8 clouds, 6' swells, 23 kt winds from 85 deg - location : 22 39.911 N, 158 0.388 W - files : run.005, pro3h180.bin * note : MPAK wasn't "blinking" when brought on deck Checking FRRF : 1446 - time OK - voltage = 14.28V - downloading data : pro3h180.bin Checking MPAK : 1514 - voltage = 12.1V - downloading data : run.005 * The battery appears to have suddenly discharged during the upcast. NOTE: The MPAK battery doesn't appear to be able to hold a charge. Recommend swapping out the core. The charger is also exhibiting some peculiarities. First the yellow light comes on, then after only a couple of minutes, the green light comes on. I would think that the green light shouldn't come on for a couple of hours.