HOT-177 Optics Readme file NOTE: FRRF & WetStar Fluorometer back from calibration. 1/23/06, 0914 - Charging FRRF & MPAK batteries - Changed all tubing - Replaced SeaCat batteries - Changed TSRB batteries - Programed TSRB 1/24/06 Checking FRRF : 900 - voltage = 14.63V - time OK Checking MPAK - reset time - voltage = 13.1V - re-formatted flashcard - ready to calibrate Calibrating AC-9 - temp = 22 C - files : 06jan24a1.dat, 06jan24c1.dat, run.002 Checking MPAK : 822 - time OK - voltage = 12.7V - ready to sample 1/26/06 Attepted a dark cast. Unable to start the FRRF! Aborted the cast. Checking the FRRF indicated that all stored protocols were erased when it was sent back for calibration. Using the following boot protocol parameters: Boot protocol #2 6. 60,000 7. 1 8. 100 9. 4 A. 0 B. Enable C. 20 D. 8 E. 61 F. 10 G. 16 *autorange mode H. Disable I. Disable J. Active K. Active L. Enable M. 93 N. 35 MPAK/FRRF cast - in:1309, down:1317, up:1348, out:1415 - depth : 250m using ship's trawl winch - rate : ~10 m/min up/down - weather : 20 kt winds from 97 deg, 10-12' swells - location : 22 45.008 N, 158 4.780 W - files : run.006 pro1h177.bin MPAK/FRRF filter cast - in:1419, down:1428, up:1457, out:1624 - depth : 250m using ship's trawl winch - rate : ~10 m/min up/down - weather : 20 kt winds from 97 deg, 10-12' swells - location : 22 45.008 N, 158 4.780 W - files : run.007 pro1h177.bin - note : lost 'A' filter 1/27/06 Downloading FRRF data : pro1h177.bin Downloading MPAK data : run.002, run.006, run.007 Note: The TSRB was not able to be recovered. Below is an excerpt from Tom Gregory's Chief Scientist Cruise Report: > We recovered the gas array at around 0200 this morning with all N2 > samples intact. The TSRB had been deployed with this array. > Regrettably, the TSRB was no longer connected to the gas array when > recovered. It must have broken free at some unknown time during the > deployment; the cause appears to have been hardware failure. We spent > several hours early this morning (while still dark) searching for > visual recognition of the strobe and/or the RDF signal but failed to > locate the instrument. The gas array drifted 10 nmi. in 24 hours. The > TSRB, once free of the array, would have rapidly drifted away from the > array.