HOT-146 Optics Readme file NOTE: Returned MPAK to Wetlabs between cruises for evaluation. Was not returned in time. All AC-9/Seacat sensors removed from "cage". 3/27/03 Recharging FRRF battery pack : 1050a - 4:42a (3/28/03) 3/28/03 Checking FRRF : 449 - time OK - voltage = 14.86 V - re-formatting flashdisk 3/29/03 FRRF cast - in:1311, up:1338, out:1406 - depth : ~240m using 280m line & capstan - rate : ~10 m/min - weather : 12-13 kt winds, seastate 3+, 7/8 clouds - location : 22 43.481 N, 158 0.324 W - files : pro1h146.bin 3/30/03 FRRF cast - in:308, up:334, out:402 (night) - depth : ~240m using 280m line & capstan - rate : ~10 m/min - weather : 11-12 kt winds, seastate 2+, 5 ft waves - location : 22 41.540 N, 158 0.240 W - files : pro2h146.bin Checking FRRF : 418 - time OK - voltage = 14.52 V - downloading data: pro1h146.bin, pro2h146.bin Re-terminated FRRF Deck Box power cable due to various power problems. PRR cast : On the downcast, the light, tilt & spin sensors went "bad" at around 100m. Once the PRR got to the bottom, the signal recovered. Everything appeared OK on the upcast. We tried a second time, lowering at a slower rate. It seemed to do better, but still went "bad" around 120m. Was this due to a dramatic density layer? I don't know. FRRF cast - in:1254, up:1320, out:1347 - depth : ~240m using 280m line & capstan - rate : ~10 m/min - weather : 12-14 kt winds, seastate 2+, 4ft swells, 8/8 clouds, raining - location : 22 45.114 N, 158 0.032 W - files : pro3h146.bin Checking FRRF : - time OK - voltage = 14.42 V - total space used = 7228478 words - downloading data: pro3h146.bin Re-termination seemed to work OK!